Errors if 0s Expected
To display the errors if "0" is expected, but "1" received.
•
Errors if 1s Expected
To display the errors if "1" is expected, but "0" received.
You can set the
Timing Unit
for the display of the relative time of the
measurement points on the Parameters tab. Choose between
Unit
Interval
and
Seconds
to select the unit.
You can select the number of
Decimal Places
to be displayed in the table.
Eye Diagram
Eye Diagram - Concepts
The Serial BERT provides quick design analysis with the Eye Diagram
capability. The Eye Diagram allows a quick check for the DUT’s signal
output, and determines the signal quality. Due to the higher sampling
depth of a BERT value, the eye contour lines display the measured eye
at a deeper BER level, for accurate results.
The Eye Diagram feature provides the following measurement
capabilities:
•
Eye Diagram Waveform
•
BER Contour Lines
•
Automatic Measurement for the most relevant eye parameters:-
– Eye Height
– Eye Width
– Jitter P-P
– Jitter RMS
– Cross Voltage
•
Markers
•
Eye Mask
Advanced Analysis
5
Agilent J-BERT N4903 High-Performance Serial BERT
249
Grid
Table Number Format
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...