Functional
T
est
4339B
Handler
Interface
Functional
T
est
The
4339B's
handler
interface
function
is
tested
using
the
built-in
selftest
and
the
handler
interface
tester
.
T
est
Equipment
Description
Recommended
Model
Handler
Interface
T
ester
04339-65007
Procedure
Initial
Setup
.
1.
Turn
the
4339B
OFF
.
2.
Set
the
LED
PW
switch
on
the
Handler
Interface
T
ester
to
OFF,
and
set
the
IN1,
IN2,
IN3,
and
IN4
switches
to
0.
3.
Set
up
the
equipment
as
shown
in
Figure
9-11
Figure
9-11.
Handler
Interface
Functional
T
est
Setup
4.
Turn
the
4339B
ON.
5.
Reset
the
4339B
using
the
following
procedure
.
a.
Press
to
display
the
system
reset
menu.
b.
Select
Yes
using
or
and
press
.
K
ey
Lock
Function
T
est.
6.
Set
the
IN4
switch
on
the
Handler
Interface
T
ester
to
1.
7.
Conrm
that
the
all
keys
on
the
4339B's
front
panel
are
locked
out.
8.
Set
the
IN4
switch
on
the
Handler
Interface
T
ester
to
0.
External
Trigger
Function
T
est.
9.
Press
to
set
the
trigger
mode
to
External.
10.
Press
and
select
I
to
set
the
measurement
mode
to
current
(I).
9-22
Maintenance
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...