![Agilent Technologies 4339B Operation Manual Download Page 127](http://html.mh-extra.com/html/agilent-technologies/4339b/4339b_operation-manual_2867889127.webp)
4339B
Getting
Started
T
o
Use
the
Comparator
Function
The
following
commands
control
the
comparator
function:
:CALC:LIM:CLE
:CALC:LIM:FAIL?
:CALC:LIM:LOW
:CALC:LIM:LOW:STAT
:CALC:LIM:UPP
:CALC:LIM:UPP:STAT
:CALC:LIM:STAT
F
or
example
,
to
set
the
limit
values
and
to
enable
the
comparator
function,
:
OUTPUT
717;":CALC:LIM:LOW
1E10"
OUTPUT
717;":CALC:LIM:UPP
1E11"
OUTPUT
717;":CALC:LIM:STAT
ON"
OUTPUT
717;":FETC?"
ENTER
717;S,D,C
:
T
o
Display
a
Deviation
Measurement
The
following
commands
display
deviation
results:
:DATA
:CALC:MATH:EXPR:NAME
:CALC:MATH:EXPR:CAT?
:CALC:MATH:STAT
F
or
example
,
to
set
the
reference
value
using
a
measurement
value
and
to
calculate
the
absolute
deviation
value
,
:
OUTPUT
717;":FETC?"
ENTER
717;s,d
OUTPUT
717;":DATA
REF,";d
OUTPUT
717;":CALC:MATH:EXPR:NAME
DEV"
OUTPUT
717;":CALC:MATH:STAT
ON"
:
T
o
Set
the
Oset-Error
Canceling
Function
The
following
command
set
the
oset-error
canceling
function
ON
or
OFF:
:CAL:AUTO
F
or
example
,
to
select
the
oset-error
canceling
function
OFF
,
OUTPUT
717;":CAL:AUTO
OFF"
T
o
W
ait
Until
Previously
Sent
Commands
are
Completed
The
following
commands
make
the
4339B
wait
until
the
previously
sent
commands
are
completed:
*OPC
*OPC?
*WAI
Remote
Operation
4-9
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...