Agilent 4339B High Resistance Meter Operation Manual
Manual Change
Agilent Part No. N/A
Jun 2009
Change 1
Add TAR and change the Measurement Uncertainty value of Source Voltage Accuracy Test (Page 9-19) to
the following.
Source
Voltage
Setting
Test Limits
Test Result
1
Measurement
Uncertainty
TAR
0 V
±0.10 V
V
± 0.00021V
485
10 V
±0.12 V
V
± 0.00027V
433
25 V
±0.14 V
V
± 0.00054V
257
50 V
±0.18 V
V
± 0.00096V
188
100 V
±0.26 V
V
± 0.00185V
141
200 V
±0.42 V
V
± 0.0037 V
115
201 V
±0.82 V
V
± 0.0040 V
206
250 V
±0.90 V
V
± 0.0053 V
170
500 V
±1.3 V
V
± 0.0097 V
133
1000 V
±2.1 V
V
± 0.0188 V
112
Change 2
Add TAR and change the Measurement Uncertainty value of Ammeter Offset Voltage Test (Page 9-19) to
the following.
Test Limits
Test Result
1
Measurement
Uncertainty
TAR
±0.5 mV
mV
± 0.0083 mV
60.27
Change 3
Add TAR and change the Measurement Uncertainty value of Ammeter Input Resistance Test (Page 9-19) to
the following.
Test Limits
Test Result
1
Measurement
Uncertainty
TAR
±50 ohm
ohm
± 4.37 ohm
11.5
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...