Getting
Started
4339B
F
or
example
,
to
wait
until
the
OPEN
correction
is
completed,
OUTPUT
717;":SENS:CORR:COLL
STAN1"
P
erform
OPEN
correction
OUTPUT
717;"*OPC?"
W
ait
for
OPEN
correction
completed
ENTER
717;A
T
o
Get
the
Current
Instrument
Settings
The
following
command
gets
the
current
instrument
settings:
*LRN?
F
or
example
,
DIM
A$[1000]
:
OUTPUT
717;"*LRN?"
ENTER
717;A$
A$
T
o
Save
and
Recall
Instrument
Settings
The
following
commands
save
and
recall
instrument
settings:
*SAV
*RCL
T
o
save
the
instrument
settings
to
register
no
.
1,
OUTPUT
717;"*SAV
1"
T
o
recall
the
instrument
settings
from
register
no
.
1,
OUTPUT
717;"*RCL
1"
4-10
Remote
Operation