4339B
Measuring
Insulation
Resistance
of
Capacitors
6.
P
erform
the
OPEN
correction.
a.
Remove
the
DUT
if
connected,
and
set
the
SMD
module's
electrode
contact
spacing
the
same
as
the
DUT's
length,
tightening
the
screw
to
hold
this
interval
between
the
electrodes
.
b.
Close
the
top
cover
.
c.
Select
the
Current
measurement
mode
.
Press
to
display
the
measurement
parameter
selection
menu.
Select
I
using
or
and
press
.
d.
Apply
the
source
voltage
by
pressing
.
e.
W
ait
until
the
current
has
stabilized
to
within
0.5
pA.
(Displayed
I
values
do
not
change
by
more
than
0.5
pA.)
f.
Press
to
display
the
OPEN
correction
menu.
g.
Select
OpenMeas
using
or
and
press
.
h.
After
the
OPEN
correction
is
nished,
press
to
turn
the
source
voltage
OFF
.
7.
Clip
the
capacitor
to
the
16339A
(see
Figure
6-2),
and
close
the
top
cover
.
Figure
6-2.
Chip
Capacitor
Binding
8.
Select
Resistance
as
the
measurement
parameter
.
Press
to
display
the
measurement
parameter
selection
menu.
Select
R
using
or
and
press
.
Application
Measurement
6-3
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...