TRIGger
Subsystem
4339B
:TRIGger[:SEQuence1]:TIMer
<numeric
value>[MS j S]
Sets
the
period
of
the
internal
periodic
signal
source
.
<numeric
value>
is
,
numeric
10
ms
to
999
s
The
range
and
resolution
are
as
follows:
Range
Resolution
10
ms
Interval
Time
<
10
s
1
ms
10
s
Interval
Time
<
100
s
10
ms
100
s
Interval
Time
999
s
100
ms
Query
response
is
a
numerical
value
in
<NR3>
format.
When
the
*RST
command
is
executes
,
period
is
sets
as
follows
.
Measurement
time
mode
P
eriod
SHORT
10
ms
MED
30
ms
LONG
400
ms
5-36
GPIB
Reference