Measuring
Resistivity
of
Insulation
Materials
4339B
11.
Measuring
the
volume
resistivity
.
a.
Turn
the
V
olume/Surface
selector
of
the
16008B
to
\V
olume".
b.
Press
.
Select
Rv
using
or
and
press
.
c.
Press
.
This
starts
the
volume
resistivity
measurement
sequence
.
Measured
volume
resistivity
value
will
be
displayed
after
60
seconds
charge
time
.
Note
If
you
want
to
abort
the
measurement
sequence
,
press
.
12.
Measuring
surface
resistivity
.
a.
Turn
the
V
olume/Surface
selector
to
\Surface".
The
4339B
id
automatically
set
to
Rs
measurement
mode
.
b.
Press
.
This
starts
the
surface
resistivity
measurement
sequence
.
Measured
surface
resistivity
value
will
be
displayed
after
60
seconds
charge
time
.
Note
If
you
want
to
abort
the
measurement
sequence
,
press
.
13.
T
o
exit
the
measurement
sequence
,
Press
.
Select
Off
using
or
and
press
.
14.
W
ait
until
the
High
V
oltage
indicator
is
OFF
.
Then
remove
the
DUT
from
the
test
xture
.
6-8
Application
Measurement
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...