Measuring
Resistivity
of
Insulation
Materials
4339B
4.
P
erform
calibration.
Press
.
Select
ExecCal
using
or
and
press
.
5.
P
erform
the
OPEN
correction.
a.
Turn
the
load
knob
counterclockwise
until
the
upper
electrode
does
not
move
.
b.
Close
the
top
cover
.
c.
Select
the
Current
measurement
mode
.
Press
to
display
the
measurement
parameter
selection
menu.
Select
I
using
or
and
press
.
d.
Apply
the
source
voltage
by
pressing
.
e.
W
ait
until
the
current
has
stabled
to
within
0.5
pA.
(Displayed
I
value
do
not
change
by
more
than
0.5
pA.)
f.
Press
to
display
the
OPEN
correction
menu.
g.
Select
OpenMeas
using
or
and
press
.
h.
After
the
OPEN
correction
is
nished,
press
to
turn
the
source
voltage
OFF
.
6.
Place
the
DUT
on
the
electrode
of
resistivity
cell.
Set
the
upper
electrode
into
position.
Figure
6-4.
Resistivity
Cell
Setup
6-6
Application
Measurement
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...