4339B
Measuring
Resistivity
of
Insulation
Materials
7.
Turn
the
load
knob
to
apply
pressure
by
the
upper
and
lower
electrodes
on
the
DUT
,
and
then
close
the
top
cover
.
Turning
load
knob
applies
load
pressure
on
the
DUT
of
up
to
approximately
10
kg.
If
you
do
not
need
to
apply
a
specic
pressure
,
DUT
must
at
least
rmly
held
by
the
upper
and
lower
electrodes
.
8.
Set
the
resistivity
cell
parameter
a.
Press
.
b.
Enter
DUT
thickness
.
Select
Thickenss
using
or
and
press
.
Enter
the
thickness
value
,
for
example
0.13
(mm),
and
press
.
c.
Select
the
resistivity
cell
size
.
Select
ResCell
using
or
and
press
.
Select
50
(mm)
using
or
,
and
press
twice
.
d.
Select
Exit
and
press
to
exit.
9.
Set
the
measurement
sequence
conguration.
Press
.
Select
Charge
and
press
.
Enter
the
charge
time
,
for
example
60
(s),
and
press
twice
.
10.
Select
the
single
mode
as
the
measurement
sequence
mode
.
Press
.
Select
Single
using
or
and
press
.
Application
Measurement
6-7
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...