P
erformance
T
ests
4339B
Current
Measurement
A
ccuracy
T
est
The
4339B
measures
the
current
through
the
calibrated
RC
Box,
and
the
measured
values
are
compared
with
the
current
values
calculated
from
the
RC
box's
calibration
value
and
the
voltage
across
the
standard.
Specication
Basic
Measurement
A
ccuracy:
6
0.4
%
(See
Chapter
8
General
Information
for
details
.)
T
est
Equipment
Description
Recommended
Model
Multimeter
3458A
RC
Box
16340A
RC
Box
A
dapter
PN
04339-65005
Dual
Banana-BNC(m)
Cable
PN
11001-60001
Procedure
1.
Record
the
16340A
RC
Box
calibration
values
on
the
calculation
sheet.
2.
Reset
the
4339B
using
the
following
procedure:
a.
Press
to
display
the
system
reset
menu.
b.
Select
Yes
using
or
and
press
.
3.
Set
the
Oset-error
Canceling
to
ON
using
the
following
procedure:
a.
Press
to
display
the
conguration
setting
menu.
b.
Select
Offset
using
or
and
press
.
c.
Select
On
using
or
and
press
.
d.
Select
Exit
and
press
to
exit
the
conguration
setting
menu.
4.
Set
up
the
equipment
as
shown
in
Figure
9-4.
9-8
Maintenance
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...