SENSe
Subsystem
4339B
[:SENSe]:CONT
act:D
A
T
A?
Returns
the
capacitance
data
at
dut
mesurement.
The
contact
check
function
uses
this
capacitance
data.
Refer
to
\Reading
the
Limit
Data
of
Contact
Check
Function
"
in
Chapter
2
for
detail.
(query
only)
Query
response
is
a
numeric
value
in
<NR3>
format.
The
data
is
incorrect
if
the
measurement
range
was
changed
after
dut
measurement.
Note
The
accuracy
of
the
capacitance
data
is
not
guaranteed.
[:SENSe]:CONT
act:LIMit?
Returns
the
contact
check
limit
data.
Refer
to
\Reading
the
Limit
Data
of
Contact
Check
Function
"
in
Chapter
2
for
detail.
(query
only)
Query
response
is
a
numeric
value
in
<NR3>
format.
[:SENSe]:CONT
act:OFFSet
<numeric
value>
Set
or
queries
an
oset
of
contact
check
limit
data.
<numeric
value>
is
,
numeric
0
to
75
[pF]
[:SENSe]:CONT
act:VERify
f
ON
j
OFF
j
1
j
0
g
Sets
or
queries
if
the
contact
check
function
is
enabled.
Where
,
ON
or
1
Enables
the
contact
check
function.
OFF
or
0
Disables
the
contact
check
function.
The
reference
data
for
the
contact
check
is
obtained
by
an
OPEN
correction.
An
OPEN
correction
is
required
before
the
contact
check
turns
on.
Query
response
is
1
or
0 .
[:SENSe]:CORRection:COLLect[:A
CQuire]
OFFset
Measures
and
stores
the
OPEN
correction
data.
(No
query)
This
command
sets
[SENSe:]:CORRection[:STATe]
ON ,
which
enables
the
correction
function.
5-26
GPIB
Reference
Summary of Contents for 4339B
Page 10: ......
Page 18: ... ᄌᦝ 0123 45 6789 8 A B C ᄌᦝ 3 DE FG H FG IJ B C K 9 C Copyright 2007 Agilent Technologies ...
Page 20: ......
Page 21: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 30: ......
Page 44: ......
Page 55: ...4339B Initial Inspection Figure 1 1 Power Cable Supplied Getting Started 1 11 ...
Page 212: ......
Page 220: ......
Page 230: ......
Page 256: ......
Page 268: ...Procedure 4339B Figure B 5 A1 Main Board B 8 Handler Interface Installation ...