Troubleshooting and Test Functions
4
1100 Series FD Reference Manual
97
Wavelength Calibration Failed
This message may show up during a wavelength calibration.
If the expected deviation is larger than the specified wavelength accuracy, the
message “Wavelength
Calibration Failed
” is displayed and the instrument
stays in a
Not Ready
condition.
Probable Causes
•
Flash lamp not ignited or position not correct.
•
Cell position not correct.
•
Solvent in the cell not clean or air bubble in the cell.
•
monochromator assembly position not correct (after replacement).
Suggested Actions
✔
Check the flash lamp image and position. See also
✔
Check the cell position.
✔
Flush the flow cell.
✔
Reset monochromator settings and re-run the wavelength calibration.