290
1100 Series FD Reference Manual
8
Theory of Operation
Setting the 8-bit Configuration Switch
The 8-bit configuration switch is located next to the GPIB connector. Switch
settings provide configuration parameters for GPIB address, serial
communication protocol and instrument specific initialization procedures.
Switches 1 and 2 define which set of parameters (for example, for GPIB,
RS-232C, and so on) will be changed. Once the change has been completed, the
instrument must be powered up again in order to store the values in the
non-volatile memory.
Figure 100
8-bit Configuration Switch
Table 64
8-bit Configuration Switch
Mode Select
1
2
3
4
5
6
7
8
GPIB
0
0
GPIB Address
RS-232C
0
1
Baud rate
Data
Bits
Parity
Reserved
1
0
Reserved
TEST/BOOT
1
1
RSVD
SYS
RSVD
RSVD
FC
Factory setting is
shown for the
fluorescence detector