Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
177
2
Remove the UV protector.
3
Locate the three slits (from left to right: PMT, EM, EX).
4
Use a pair of tweezers to replace the slit.
5
Put slight pressure on to the spring part of the slit.
Next Steps:
•
Replace the UV protector.
•
Replace the front foam.
•
Replace the covers as described in
“Replacing the Foams and Covers"
•
Re-install the flow cell and the front panel
•
Re-install the detector in the stack.
•
Turn the lamp on.
•
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in chapter