Troubleshooting and Test Functions
4
1100 Series FD Reference Manual
119
Raman ASTM Signal-to-Noise Test
This test verifies the Raman ASTM signal-to-noise.
Table 21
Raman Signal-to-Noise Test Conditions
Duration
approximately 23 minutes
Report Style (Agilent ChemStation)
PerfNoise
Noise Determination
5 to 20 minutes
Solvent
LC grade water, degassed
Flow rate
0.5 - 1 ml/min
Figure 47
Raman ASTM Signal-to-Noise Test
Dark current at
350/450 nm
and PMT=12
ASTM Noise and drift at 350/397 nm, PMT=12 and
Responsetime=8 s taken from 5 to 20 minutes