1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Removing the Covers
When required
For all repairs inside the detector
Tools required
Screwdriver POZI 1 PT3
Parts required
Depends on the work inside and the following procedures
C A U T I O N
There will be a risk of damaging hardware due to overheating when operating the
instrument without covers.
Preparations for this procedure:
•
Turn off the detector.
•
Disconnect the power cable.
•
Remove the flow cell.
•
Remove detector from stack and place it on the
workbench.
1
Press the release buttons and remove the front cover.
Unclip and remove the waste funnel assemly.