76
1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
Detector Status Indicator
The detector status indicator indicates one of four possible detector
conditions:
•
When the status indicator is
OFF
(and power switch light is on), the
detector is in a
prerun
condition, and is ready to begin an analysis.
•
A
green
status indicator, indicates the detector is performing an analysis
(
run
mode).
•
A
yellow
indicator indicates a
not-ready
condition. The detector is in a
not-ready state when it is waiting for a specific condition to be reached or
completed (for example, immediately after changing a setpoint), or while a
self-test procedure is running.
•
An
error
condition is indicated when the status indicator is
red
. An error
condition indicates the detector has detected an internal problem which
affects correct operation of the detector. Usually, an error condition
requires attention (e.g. leak, defective internal components). An error
condition always interrupts the analysis.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Page 12: ...12 1100 Series FD Reference Manual ...
Page 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Page 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Page 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Page 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Page 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Page 345: ......