
80
1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
Remote Time-out
A not-ready condition is still present on the remote input.
When an analysis is started, the system expects all not-ready conditions (e.g. a
not-ready condition during detector balance) to switch to run conditions
within one minute of starting the analysis. If a not-ready condition is still
present on the remote line after one minute the error message is generated.
Probable Causes
•
Not-ready condition in one of the instruments connected to the remote line.
•
Defective remote cable.
•
Defective components in the instrument showing the not-ready condition.
Suggested Actions
✔
Ensure the instrument showing the not-ready condition is installed
correctly, and is set up correctly for analysis.
✔
Exchange the remote cable.
✔
Check the instrument for defects (refer to the instrument’s reference
documentation).
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Page 12: ...12 1100 Series FD Reference Manual ...
Page 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Page 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Page 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Page 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Page 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Page 345: ......