
294
1100 Series FD Reference Manual
8
Theory of Operation
Stay-Resident Settings
Firmware update procedures may require this mode in case of firmware
loading errors.
Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident part, that is, it is not operable as
a detector. It only uses basic functions of the operating system for example, for
communication.
To return to normal operation, set switches back to your GPIB or RS-232C
configuration settings.
Table 71
Stay Resident Settings
Mode Select
1
2
3
4
5
6
7
8
TEST/BOOT
1
1
0
0
1
0
0
0
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Page 12: ...12 1100 Series FD Reference Manual ...
Page 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Page 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Page 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Page 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Page 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Page 345: ......