1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
The Wavelength Calibration Procedure
Steps
1
Preparation of the Glycogen Calibration Sample.
2
Preparation of the Flow Cell.
3
Wavelength Calibration.
4
“Check of Raman Band of Water"
Preparation of the Glycogen Calibration Sample
1
To prepare 10 ml of the calibration solution you have to use 10 mg of the
Glycogen sample (a tolerance of
±
20% is not critical).
2
Fill the prepared amount into a suitable bottle/vial.
3
Fill 10 ml of distilled water into the vial and shake.
4
Wait 5 minutes and shake again. After 10 minutes the solution is ready.
When required
If application requires, or see
Tools required
Laboratory balance
Parts required
Glycogen Calibration Sample, Syringe, needle, sample filter and PEEK fitting
from the Accessory Kit