200
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Replacing the Detector’s Firmware
The installation of new firmware is required:
•
if new version solves problems of currently installed version, or
•
if after exchange of the main board (FLM) the version on board is older
than the previously installed one.
To upgrade the detector’s firmware the following steps have to be taken:
1
Load the firmware into the detector, see help system of your user interface.
2
If FLM board was replaced, reenter the serial number information of the
module through the user interface, see
“Entering the Serial Number using
on page 161 or
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Page 12: ...12 1100 Series FD Reference Manual ...
Page 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Page 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Page 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Page 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Page 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Page 345: ......