Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
167
5
Use a flat screw driver to lift the leak base from the base.
6
Remove the leak base completely.
7
Replace the leak pan into the cabinet bottom. Press the
leak pan completely down into its recess.
8
Replace the leak sensor using your left hand and
reconnect the connector.
Next Steps:
•
Replace the covers as described in
“Replacing the Foams and Covers"
•
Re-install the flow cell and the front panel
•
Re-install the detector in the stack.