112
1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
Wavelength Calibration
1
From the user interface start the FLD Wavelength Calibration (see
Agilent ChemStation:
Diagnosis - Maintenance - FLD Calibration
Control Module:
System - Tests - FLD - Calibrate
.
2
If a deviation is displayed, press Adjust and OK. The history table will be
updated.
N O T E
If the wavelength calibration process fails, refer to
“Wavelength Calibration Failed"
N O T E
To look at the history table (ChemStation) start a wavelength calibration and abort
immediately. No changes are made to the calibration at this time.
N O T E
Rinse the flow cell with pure water at a minimum of 1.5 ml/min to get rid of the Glycogen
from the cell and the capillaries. When organic solvent is sequentially applied (without
rinsing), a blockage of capillaries may occur.