
Troubleshooting and Test Functions
4
1100 Series FD Reference Manual
99
Flow Cell Removed
The detector has an automatic cell recognition system. When the flow cell is
removed, the lamp is turned off and a NOT READY condition exists. If the flow
cell is removed during an analysis, a SHUT DOWN is generated.
Probable Causes
•
Flow cell has been removed during analysis.
Suggested Actions
✔
Insert flow cell and turn on the lamp.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Page 12: ...12 1100 Series FD Reference Manual ...
Page 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Page 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Page 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Page 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Page 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Page 345: ......