5
Repairing the Fluorescence Detector
Assembling the Main Cover
When required
If cover is broken
Tools required
None
Parts required
Plastics kit 5062-8582 (includes base, top, left and right)
N O T E
The plastics kit contains all parts, but it is not assembled.
WA R N I N G
If you mistakenly insert the left or right side in the opposite position, you may not be
able to remove that side from the top part.
1
Place the top part on the bench and insert the left and right
sides into the top part.
2
Replace the cover.
Next steps:
•
Replace the detector in the stack and reconnect the cables and capillaries.
•
Turn on the detector.