1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
2
Disconnect the capillaries from the flow cell.
3
Unscrew the thumb screws and pull the flow cell out of
the compartment.
Note:
The label attached to the flow cell provides
information on part number, cell volume and
maximum pressure. The cell type will be
automatically detected.
There are no parts that can be replaced on the flow
cell. If defective (leaky) the flow cell has to be
replaced completely.
4
Insert the flow cell and tighten the thumb screws.
Reconnect the capillaries to the flow cell. DO NOT install
the inlet capillary to the outlet connection of the flow cell.
This will result in poor performance or damage.