186
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
10
Replace the lamp housing cover. Insert the optical unit
into its location and carefully insert the rear foam (A).
Then reconnect the cables.
Next Steps:
•
Replace the covers as described in “Replacing the Foam
and the Top Cover” on page149.
•
Re-install the detector in the stack.
•
Reset the lamp counter as described in the user
interface documentation.
•
Turn the lamp on.
•
Perform a wavelength verification to check the correct
positioning of the replaced assemblies, as described in
chapter
“Wavelength Verification and Calibration"
A
B
A
B
A
EM
EX
Rear
foam