Theory of Operation
8
1100 Series FD Reference Manual
291
In the non-volatile memory, the parameters are kept, regardless of whether
you turn the instrument off and on again. They will be kept until the same set
of parameters is changed and the power is reset. All other previously stored
configuration settings will still remain in the non-volatile memory.
In this way, you can store more than one set of parameters using the same
8-bit configuration switch twice, for example, for both GPIB and RS-232C.
GPIB Default Addresses
If you just want to change the GPIB address and need a detailed procedure,
refer to the
Installing Your Agilent ChemStation System
handbook.
Default GPIB address is set to the following addresses:
where 0 means that the switch is down and 1 means that the switch is up.
Table 65
Default Addresses for Agilent Series 1100 Modules
Module
Address
Binary Address
Pump
22
0 0 0 1 0 1 1 0
FLD
23
0 0 0 1 0 1 1 1
VWD
24
0 0 0 1 1 0 0 0
Agilent 8453A
25
0 0 0 1 1 1 0 1
DAD / MWD
26
0 0 0 1 1 0 1 0
Column compartment
27
0 0 0 1 1 0 1 1
Autosampler
28
0 0 0 1 1 1 0 0
RID
29
0 0 0 1 1 1 0 1
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Page 12: ...12 1100 Series FD Reference Manual ...
Page 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Page 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Page 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Page 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Page 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Page 345: ......