1100 Series FD Reference Manual
2
First Steps with the Fluorescence Detector
Previously, only diode array detectors and mass spectrometric detectors could
deliver spectral information on-line to confirm peak identity as assigned by
retention time.
Now, fluorescence detectors provide an additional tool for automated peak
confirmation and purity control. No additional run is necessary after the
quantitative analysis.
During method development, fluorescence excitation and emission spectra are
collected from reference standards and entered into a library-at the choice of
the method developer. All spectral data from unknown samples can then be
compared automatically with library data. Table 3 illustrates this principle
using a PNA analysis. The match factor given in the report for each peak
indicates the degree of similarity between the reference spectrum and the
spectra from a peak. A match factor of 1,000 means identical spectra.
Figure 17
Simultaneous multi wavelength detection for PNA-analysis
The upper trace was
received with
traditional wavelength
switching.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Page 12: ...12 1100 Series FD Reference Manual ...
Page 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Page 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Page 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Page 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Page 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Page 345: ......