Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
189
6
Bring the top of the assembly into the position shown
below and fix the 3 screws.
7
Assure the UV protector cover is installed. Replace the
foam on the optical unit.
Next Steps:
•
Reconnect the encoder cable of the EX-monochromator to FLF board.
•
Reinstall the optical unit as described in
•
Enter the replacement in the maintenance logbook.
•
Reset the monochromator settings to default using the diagnose functions.
•
In this case “
Wavelength Calibration Lost”
is displayed and the instrument stays in
Not Ready
condition until you
perform a wavelength calibration.
•
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in chapter