120
1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
Formula for the Raman ASTM S/N value
(see
for details):
The result should be
> 200
with FLF board revision A.
The result should be
> 400
with FLF board revision B and above.
Figure 48
Raman ASTM signal/noise calculation
RamanASTM
Height RamanCurrent
(
)
Height DarkCurrent
(
)
ñ
ASTMNoise
-------------------------------------------------------------------------------------------------------------------------------
=
Height Raman
Current
average between
5 and 10 minutes
Height Dark Current
average between
21 and 22 minutes
ASTMNoise (time 5 to 20 minutes)