Troubleshooting and Test Functions
4
1100 Series FD Reference Manual
79
Shutdown
An external instrument has generated a shut-down signal (through CAN or
REMOTE lines).
The detector continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
•
Leak detected in an external instrument with a remote connection to the
system.
•
Shut-down in an external instrument with a remote connection to the
system.
•
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
✔
Fix the leak in the external instrument before restarting the pump.
✔
Check external instruments for a shut-down condition.
✔
Check the degasser for an error condition. Refer to the
Reference Manual
for the Agilent 1100 Series vacuum degasser.