278
1100 Series FD Reference Manual
8
Theory of Operation
Data flow for spectral output
The data flow is shown below.
Figure 97
Data flow for spectral output
PMT raw signal
acquisition
position
reference diode
raw signal
acquisition position
dark
subtraction
dark
subtraction
normalization of
reference signal
divide
PMT / reference
collect data for
complete spectra
spectra output
(LU)
Influences
normalization to
spectra
LU
concentration
correction
ChemStation
Spline algorithm
Data Analysis
Control Module
Online Monitor