126
1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
Procedure using the Control Module
1
Load the FLD default parameter
•
peakwidth to >0.2 minutes (standard),
•
set stop time to 4.8 minutes.
2
Enter System -
Tests - FLD
.
3
Select the function
Enable test chromatogram
and press
Execute
.
The logbook shows “Simulated chromatogram enabled”.
4
Enter the
Plot
function and select
Signal A
and/or
Analog x
, set a
time
range
of 10 minutes and the
FLD signal
from -1 to 50 LU.
5
Start the run.
After the run has finished, the plot can be printed to a connected printer by
pressing the m-key and selecting
Print Plot
.
N O T E
If an Agilent ChemStation is connected, the next step will start the Agilent ChemStation
simultaneously and will overwrite the previous used Data File!
N O T E
The test chromatogram is switched off automatically at the end of the run.