168
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Replacing the EM-Condenser or Cutoff-Filter
When required
If light throughput is too low
Tools required
Screwdriver POZI 1 PT3
Hexagonal key 2.5 mm
Parts required
Condenser assembly EM 1000-1124
Cutoff filter 295 nm 5062-8530
Preparations for this procedure:
•
Turn off the detector.
•
Disconnect the power cable.
•
Remove the flow cell.
•
Remove detector from stack and place it on the
workbench.
•
Remove the covers as described in
•
Remove the flow cell.
1
Disconnect the emission and the excitation cables and
from the detector main board.
A
B
A
B
A
EM
EX