Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
135
Exchanging a Flow Cell
ü
When required
If an application needs a different type of flow cell or the flow cell is defective
(leaky).
Tools required
Two 1/4 inch wrenches for capillary connections
Parts required
Standard flow cell, 8 µl, 20 bar, G1321-60005
Cuvette for off-line measurements, 8 µl, 20 bar, G1321-60007, refer to
on page 138 for more information on usage.
N O T E
DO NOT install the inlet capillary to the outlet connection of the flow cell. This will result in
poor performance.
Preparations for this procedure:
•
Turn off the flow.
1
Press the release buttons and remove the front cover for
access to the flow cell area
.