Troubleshooting and Test Functions
4
1100 Series FD Reference Manual
105
6
Load the method WLEMTEST.M. The FLD will change into the
multi-excitation mode and scan in the range of the expected maximum of
350 nm
±
20 nm.
7
Define in
Sample Information
the file name and description.
8
Start a run. After the 1 minute run has completed, enter the Data Analysis
and evaluate the spectrum. The maximum should be found at
350 nm
±
3 nm.
Figure 34
Emission Spectrum