1100 Series FD Reference Manual
A
Safety Information
Declaration of Conformity for HOX2 Filter
We herewith inform you that the
holmium oxide glass filter (type Hoya HY-1)
meets the following specification of absorbance maxima positions:
360.8 nm – 418.5 nm – 536.4 nm (spectral bandwidth: 1 nm).
Agilent Technologies guarantees the traceability of the specified absorbance
maxima to a National Institute of Standards & Technology (NIST) holmium
oxide solution standard with a lot-to-lot tolerance of ± 0.3 nm.
The wavelength calibration filter built into the Agilent Technologies UV-visible
detectors
is made of this material and meets these specifications. It is, therefore,
suitable for wavelength calibration of these detectors within the specified
wavelength accuracy of the respective detector over its wavelength range.
November 1, 1995______________
DateQuality Manager
Table 78
Wavelength Accuracy
Product number
79853C
HP 1050 Series
± 2 nm
G1314A
Agilent 1100 Series
± 1 nm