324
1100 Series FD Reference Manual
9
Introduction to the Control Module
Lamp Characteristic
Use the F5 key (
Lamp Char
) to get the instrument profile. Use F8 key (
Take
)
to generate a new profile.
Wavelength Calibration
Use the F1 key (
Calibrate
) and then F8 key (
Execute
) to start the FLD
wavelength calibration (with Glycogen in the flow cell). If a deviation is found,
press
Adjust
.