100
1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
No Peaks
If no peaks are shown in the chromatogram, the user-interface shows the
module still in “Ready”. There is no feedback mechanism that checks whether
the lamp is on.
Probable Causes
•
defective FLL board / Trigger pack
•
defective Xenon flash lamp
•
wrong position of monochromator
•
FLF board defective
Suggested Actions
✔
Perform a
on page 115. If no profile available (very
low counts),
✔
replace FLL board / Trigger pack
✔
replace Xenon flash lamp
✔
perform a
“Check of Raman Band of Water"
on page 104 to check the
wavelength calibration.
✔
replace FLF board