5 First Operating Steps | 5.7 Working with Specific Specimen Types
ZEISS
Pressure-limiting aperture
Mid-column aperture
None (HV mode)
20 μm or 30 μm
400 μm EasyVP aperture
20 μm or 30 μm
100 μm VP aperture
750 μm
Fig. 42: 100 μm
aperture or 400 μm EasyVP aperture
Parts and Tools
400 μm EasyVP aperture (354720-1464-000)
Alternatively: 100 μm VP aperture (350700-0700-000)
Aperture removal tool (350700-0692-000)
VP aperture tool (350700-0859-000)
Prerequisite
ü
The microscope is equipped with the VP mode or EasyVP mode option.
Procedure
1. In the
SEM Controls
panel, select the
Vacuum
tab.
2. To vent the specimen chamber, click
Vent
.
3. If a lens-mounted BSD is fitted, remove the BSD
and place it in the parked position.
4. Retract all retractable detectors.