6 Care and Maintenance | 6.3 Maintenance Work
6.3 Maintenance Work
Only conduct maintenance work described in this document. All tasks of maintenance, service and
repair not described here must only be performed by an authorized ZEISS service representative.
6.3.1 Change of Consumables and Chemicals
The change of consumables and chemicals has to be performed by a ZEISS service representative
at mandatory intervals.
The times scheduled are designed for the maximum equipment performance level (i.e. 24 h per
day of permanent operation).
Interval
Component/Part
As required
Filament
Yearly or as required
Tip seal and tip O-ring of the pre-vacuum
pump
Yearly performance check
§
§
All optional detectors
Tab. 5: Schedule for the change of consumables
6.4 Care and Cleaning Work
All care and cleaning work not described here must only be performed by an authorized ZEISS ser-
vice representative.
NOTICE
Functional impairment due to dirt and moisture
Dirt, dust and moisture can impair the Microscope System’s functionality and can cause short-
circuits.
4
The ventilation slots must be unobstructed at all times.
4
Perform regular maintenance and cleaning according to the instructions in this document
and according to the instructions in the applicable documents.
4
Make sure that no cleaning liquid or moisture gets inside the Microscope System.
4
In case of damage, the affected parts of the Microscope System must be taken out of op-
eration.
6.4.1 Cleaning the Microscope
If required, use a moistened lint-free cloth to clean the outside of the microscope.
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006