9 Technical Data and Conformity | 9.4 Performance Data and Specifications | EVO 15
Parameter
Value
Ampere interrupting capacity (AIC)
Min. 10,000 A
rms
Location
Requirements
Parameter
Requirement
Installation site
Exclusively inside buildings
Recommended
room size
Min. 3.6 m × 4.0 m × 2.3 m
Service area
Min. 0.8 m at each side
Entrance
Min. 0.8 m wide
Hallways
Min. 1.0 m wide
Corners
Min. 1.2 m wide
Transport ways
Free of staircases
Installation cate-
gory
II
Floor stability
> 1000 kg/m²
Exhaust Line
If toxic chemicals or biological specimens are used an exhaust line is recommended to remove the
waste gas of the pre-vacuum pump and to transmit it to the outside.
Nitrogen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. The nitro-
gen can be taken either from a gas cylinder or from an in-house supply system.
Parameter
Requirement
Flow rate
Approx. 3 l/min for ventilation of specimen chamber with chamber
door open
Pressure
0.2–3.3 bar
Quality
4.6 with nitrogen content > 99.996 %
Connection hose
4 mm inside diameter. 10 m are delivered with the microscope.
Instrument con-
nection
Quick exchange connector. One is delivered with the microscope.
Compressed Air
Supply
Compressed air is used to operate several valves and the auto leveling system. The necessary com-
pressed air can be either generated by a compressor (part no. 345596-0000-000) or taken from a
gas cylinder or from an in-house supply system.
Environmental
Requirements
Parameter
Requirement
Electrical field
The microscope is a class A device (industrial). The microscope is de-
signed to operate in a controlled electromagnetic environment. This
means that devices with RF transmitters such as mobile phones or
DECT phones must not be used in close proximity.
Vibrations
Less than 6 μm/sec rms from 0–30 Hz
Less than 12 μm/sec rms above 30 Hz
Magnetic stray
fields
Max. 3 × 10⁻⁷ T (peak to peak) = 3 mG (peak to peak)
Acoustic noise
Less than 53 dB for frequencies up to 200 Hz
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006
161