USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 5
(3/08) Calibration
-
Page 5-29
Note:
The software will indicate when calibration is complete and
a PASS condition has been achieved.
5.
Once calibration is complete and within tolerance you can view or
print the
Report
.
6.
Click
Finish
to return to the C
ALIBRATION
M
AIN
M
ENU
.
Note:
The
Halt
button will stop calibration before it is complete.
ATTENTION:
Make sure the tiles are observed in the order requested
on the screen. During this procedure the tile being
measured is compared to the standard value stored for
that tile. If the values are different, the system adjusts
the constants, which could cause measurements to be
skewed.
Note:
Always use the same set of tiles for daily calibration that you
used for original color calibration.
DO NOT MIX TILE SETS
.
5.2.3.1
Preview Image
The image preview box
launches automatically when
Color/Trash Calibration is
selected. You can select to
display the image as is, or to
show the trash.
Figure 5-20: Preview Image
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......