Chapter 6
(3/08) System
Testing
-
Page 6-28
_
When the color/trash hand rises, the first and second
observations have been made. The module box keeps count of
the number of observations made for each cotton sample and the
number that is required for the cotton sample.
Example:
2/4
means that two of the four required repetitions have been
completed.
_
When all the color/trash tests have been made, remove the
cotton samples from the trays. If the Rd value is less than 40
(<40) or if it is greater than 87 (>87) or, if the +b value is <4 or
>18 the color test must be repeated.
Once all the repetitions are completed, the microprocessor calculates
the averages of the test results and displays them on the screen.
Note:
Trash image analysis or color data may disqualify a CT
measurement.
10
Exit Testing
If you have completed System Testing and want to exit, click the
Close
Testing
button.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......