Chapter 4
(3/08) Configuration
-
Page 4-7
3
Lot Limits
Lot Limits establish a range of legitimate property values. Lot limits
apply to the outcome for a bale in system testing. Lot limits do NOT
apply to calibration, module testing, module qualification, or
diagnostics.
When
L
OT
L
IMITS
are
ON
, a cotton sample property measured during
System Testing that falls outside the valid range will cause an error
message to be displayed. This alerts the operator that the cotton sample
is not acceptable.
When the L
OT
L
IMITS
M
ENU
is displayed, Upper and Lower Limits can be
entered for Micronaire, Rd, +b, Trash Area, Trash ID, UV, NEP, UH-
Mean Length, Uniformity Index, Short Fiber Index, Strength, Elongation,
Moisture, Maturity, SCI.
Figure 4-4: Lot Limits Menu
Once you have selected/edited the lot limits you would like used in
system testing, click
Apply
and go to the T
EST
S
ETUP
M
ENU
. Place a
check mark in the box preceding
Apply Lot Limits.
Checking this
selection will enable or disable the entire lot limit set for the test mode
selected.
Note:
If you
Close
before selecting
Apply
, your lot limit selections
will not be saved.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......