Chapter 5
(3/08) Calibration
-
Page 5-18
4.3
Check Calibration
This calibration procedure allows you to calibrate the micronaire unit for
low MIC (fine) or high MIC (course) cotton.
To access the M
ICRONAIRE
C
ALIBRATION
M
ENU
from the C
ALIBRATION
M
ENU
highlight
Micronaire
in the Module window, then highlight
Mic
Cotton
in the Calibration Type window and click
Check Calibration
.
Figure 5-9: Micronaire Calibration Screen
To perform the Mic Cotton orifice calibration:
1.
Insert the
low
mic cotton into the micronaire chamber.
2.
Click the
Start
button on the screen.
3.
Remove the
low
mic cotton.
4.
Insert the
high
mic cotton into the micronaire chamber.
5.
Click the
Start
button on the screen.
6.
Remove the
high
mic cotton.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......