Test of Pattern 2
Failed.
Restart program.
“
File
VP.CFG
Not
Found.
Copy file from the HVI directory
of another computer and copy
into the C:/HVI directory of the
computer that displayed the
error.
“
Lens Tare Voltage
Too High.
“
Small Sample (Low
Amount).
“
Large Sample (High
Amount).
“ Negative
Span
Length.
“ Could
Not
Find
Break Amount.
“ ERROR:
Break
Index Less Than 1.
“ Bad
Break
Steps.
These messages only indicate a
problem if they are persistent.
If persistent, go to Diagnostics
and select
LS Step by Step
.
Perform a
Continuous Test
,
and check
Fibrogram
Quality
to diagnose the
problem.
Fibrogram
“ Fibrogram
Has
Not
Enough Data Points
(%1, Normal=81).
1. Possible Control board
error.
2. Go to Length Motor
Diagnostics and check
motor positions. Reset
Position 0, Move to Home,
Move to Comb1 Front. If
current position value is
larger than the learned
value above it is OK. If it is
smaller, it indicates a
problem. Repeat the
procedure for Comb 2.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......