Electrical Requirements
The
HVI 1000
system operates on 220V AC
±
15% (50 to 60 Hz) and
requires a separate dedicated 15-amp circuit breaker at the facility’s
electrical load center. During normal operation, the
HVI 1000 system
draws approximately eight (8) amps; the startup current is sufficiently
high to require a larger breaker. A power cord is provided that will
connect to the
HVI 1000
system. The opposite end must be fitted with
connections compatible with the facility’s electrical system. The power
line is a single phase main with a protection switch. All the system
components are plugged into the power bar located in the Micronaire
cabinet with the exception of the vacuum system, which is plugged into
the Length/Strength cabinet power box.
3
Required Laboratory Conditions
It is recommended that the instrument be operated in a controlled
environment, in accordance with the following ASTM standards, as
follows:
ASTM D1776:
Humidity 65%
(
±
2%)
Temperature
21
°
C (
±
1
°
), 70
°
F
±
2
°
F
Cotton samples should be properly conditioned in the laboratory
environment to maintain consistent results.
To attain moisture equilibrium of a cotton sample, a conditioning time of
at least 24 hours is required, 48 hours is preferred. Refer to the
applications handbook for further details.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......